Gas Metering
Coriolis Meters on Gas: Sizing by Tube Mach Number
On gas, a Coriolis meter is limited by the gas velocity in its measuring tubes relative to the speed of sound, and the line or flange size does not change that limit.
On gas, the limit is the speed of sound
Coriolis meters are best known in liquid service, but they are also used for gas custody transfer, where AGA Report No. 11 and OIML R137-1 are the usual references. On gas, what limits a Coriolis meter is the velocity of the gas inside its measuring tubes compared with the speed of sound in that gas: the Mach number.
Emerson's data sheet for the Micro Motion ELITE series (PS-00374 Rev AO) gives the gas performance as a mass flow accuracy of ±0.25 % of rate and a repeatability of 0.20 % of rate, with mass flow linearity of ±0.05 % of rate up to Mach 0.2 (p.5). The linearity limit is tied to a Mach number, not to a flow rate or a line size.
Sizing to Mach 0.2 and Mach 0.3
The same data sheet sets two Mach numbers for sizing. Mach 0.2 gives the typical nominal gas rate, and Mach 0.3 gives the maximum recommended rate. Above 0.3, it says, most gas flows become compressible and pressure drop may rise significantly, whatever device is doing the measuring (p.8).
For dual-tube sensors the rate in kg/h follows from ṁ = %M × ρ × VOS × (π/4 × D² × 2) × 3600. Here %M is the chosen Mach number, ρ the gas density at operating conditions, VOS the speed of sound in the gas and D the inside diameter of one measuring tube; the factor 2 counts the two tubes. Neither the line size nor the flange size appears in it: at a given Mach number, density and speed of sound, the gas capacity is set by the measuring tube diameter of the sensor model.
The data sheet also notes that the gas maximum can never exceed the meter's maximum liquid rate. Whichever of the two is lower applies (p.8).
The manufacturer's worked example
Emerson works one case through. A CMF300M measures natural gas with a molecular weight of 19.5 at 16 °C and 34.47 barg, where the gas density is 24 kg/m³ and the speed of sound 430 m/s. The CMF300M measuring tube ID is 44.7 mm. At Mach 0.3 the data sheet arrives at 34,988 kg/h and calls it the maximum recommended rate for that meter on that gas at those conditions (p.8).
Because density sits directly in the formula, the mass rate at a given Mach number falls with line pressure. A meter that is comfortable at full line pressure can reach Mach 0.3 during a low-pressure period at a flow it handled easily before.
Emerson recommends sizing every gas application in its Flow Measurement Sizing & Selection Tool, which reports both the actual velocity and the sonic velocity for each flow rate and meter size (p.8). The hand formula is good for a first check. The sizing record should carry the tool's output, at the normal and at the lowest operating pressure.
The pressure effect
A Coriolis meter on gas usually runs well away from the pressure it was calibrated at, and its readings shift slightly with pressure. For the CMF300M of the worked example, the ELITE data sheet gives typical values of −0.009 % of rate per bar on mass flow and +0.0029 kg/m³ per bar on density, and says the effect can be corrected by a live pressure input or a fixed meter factor. Where the meter's own calibration sheet gives a pressure compensation coefficient, that coefficient applies (p.18). Other models in the same table have other values.
The arithmetic is short. If line pressure sits 8 bar away from the calibration pressure, 8 × 0.009 = 0.072 % of rate, a little over a quarter of the ±0.25 % accuracy figure. Left uncorrected, that is not scatter. It is a bias, and it moves with the line pressure.