Hydrocarbon Dew Point: Why It Is Measured at Reduced Pressure
A hydrocarbon dew point means something only together with the pressure it was measured at and a sample system kept warmer than the gas's own dew point.
A dew point without a pressure is incomplete
Water dew point and hydrocarbon dew point are both the temperature at which the first liquid forms from the gas, but they do not behave the same way with pressure. A hydrocarbon dew point does not simply rise with pressure. Natural gas has a phase envelope. Following its dew point curve down from high pressure, the hydrocarbon dew point first rises as pressure falls (the retrograde region), peaks, and then falls with pressure. The peak is the cricondentherm, the highest temperature on the envelope: above it, no hydrocarbon liquid forms at any pressure.
That is why a hydrocarbon dew point needs its pressure next to it. Michell's Condumax II manual (97081 Issue 30.7) describes the practice: “established business practice at custody transfer is to measure water dew point at full line whilst HC dew point is determined at intermediate pressure, commonly 27 barg (400 psig), the cricondentherm condition at which HC dew point temperature will be highest on the retrograde phase envelope” (p.14). The manual illustrates this with a typical phase envelope for Northern Europe natural gas, so 27 barg is a common value, not a fixed one. The dew point at the cricondentherm is the figure that tells whether liquid can drop out anywhere downstream.
How a chilled-mirror cell sees condensate
The Condumax II cools a small optical surface, a shallow, abraded, conical depression, with a Peltier cooler that can take it more than 50 °C below the analyzer's own temperature. A collimated light beam is scattered by that surface. When hydrocarbon condensate forms on it, the scattered light intensity falls, and when it reaches a user-set trip point the surface temperature is recorded as the dew point (pp.1 and 3).
Each measurement traps a fixed volume of sample in the cell: flow stops at the start of the cycle and resumes once the dew point is found (p.1). The minimum cycle time is 10 minutes, and under ideal conditions the recovery phase takes about 8 minutes. High dew points, within 10 °C of saturation at the sampling system temperature, may need longer cycles so that residue does not build up on the surface. The water dew point, from a separate sensor, is measured continuously (p.3).
The hydrocarbon sensor is calibrated at three points across its range with a certified mixture of 10 mol % n-butane in nitrogen, and the reference dew point of that mixture at each pressure is calculated with the Peng-Robinson equation of state (p.3). ISO/TR 12148 covers the calibration of chilled-mirror hydrocarbon dew point instruments, and ISO 6327 the measurement of water dew point with cooled-surface hygrometers.
Other instruments read differently
Other chilled-mirror hydrocarbon dew point analyzers use different optical arrangements, so their detected signal need not move at the dew point the way the Condumax II's scattered light does. Each reading, and any trip point or alarm set on it, is interpreted against that instrument's own manufacturer description.
The sample system decides what the analyzer sees
A dew point analyzer can only report the gas that reaches it. If the sample cools below its hydrocarbon dew point anywhere on the way, the heavier components condense in the tubing, the gas arriving at the cell is leaner, and the analyzer reports a dew point lower than the line's. The Condumax II manual therefore requires the sample conditioning system to stay above the highest expected water and hydrocarbon dew point, at both the process sample pressure and the analysis pressure (p.14).
Reducing the sample to analysis pressure cools it by the Joule-Thomson effect, and the manual asks for that cooling to be offset by direct heating. For outdoor systems the enclosure has to sit in full shade from direct sun, with a sun canopy if necessary (p.14). Shade matters for the measurement too: the cooler's reach is counted from the analyzer's own temperature, so an enclosure heated by the sun leaves less range below it.